ACCENT DR RF
Report
- Report Number
- 2017865-2013-00826
- Event Type
- Injury
- Date Received
- January 9, 2013
- Date of Event
- August 22, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- DXY
- PMA / PMN Number
- NA
- Removal / Correction Number
- Z0429
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. (B)(4) - NO COMPLAINT WAS RECEIVED WITH RETURN OF THE DEVICE. FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT AN INTEGRATED CIRCUIT ANOMALY RESULTED IN A HIGH CURRENT DRAIN.
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD.
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
ADDITIONAL INFORMATION, IT WAS REPORTED THAT THE PULSE GENERATOR EXHIBITED EXCESSIVE BATTERY DISCHARGE. THE DEVICE WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 12481 | ACCENT DR RF | IMPLANTABLE PACEMAKER PULSE GENERATOR | DXY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | PM2210 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 82 YR | Required Intervention |