BIOGLUE SURGICAL ADHESIVE
Report
- Report Number
- 1063481-2013-00002
- Event Type
- Injury
- Date Received
- January 9, 2013
- Report Date
- June 10, 2013
- Manufacturer
- CRYOLIFE, INC.
- Product Code
- MUQ
- PMA / PMN Number
- P01003
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SP
- Reporter Occupation
- PHYSICIAN
Narratives
ACCORDING TO THE PUBLICATION BY RUIZ-ZAFRA ET AL., THE AUTHORS STATE THAT BIOGLUE CAUSES FALSE POSITIVES IN FDG-PET/CT IMAGING DUE TO FOREIGN BODY REACTIONS TO THE ADHESIVE, WHICH MAY LEAD TO UNNECESSARY SURGERY. THE PATIENT UNDERWENT A LEFT THORACOTOMY AND ATYPICAL PULMONARY RESECTION FOR SQUAMOUS CELL CARCINOMA. BIOGLUE WAS USED AS AN ADJUNCT TO STAPLED SUTURE LINE. SIX MONTH LATER, FDG-PET/CT IMAGING REVEALED THE PRESENCE OF HYPERMETABOLIC FOCI IN THE RESECTED AREA OF THE LEFT UPPER LOBE, WITH A SUVMAX OF 4.66 SUGGESTIVE OF TUMOR REOCCURRENCE. RE-THORACOTOMY WAS REQUIRED, REVEALING THE PRESENCE OF FIBROTIC TISSUE AT THE SURGICAL SITE. THE SURGEONS FOUND AN ENCAPSULATION SURROUNDING THE AREA WHERE THE BIOADHESIVE WAS APPLIED. HISTOLOGICAL STUDY SHOWED A LESION COMPATIBLE WITH FOREIGN BODY GRANULOMA. AN ARTICLE TITLED "PERSISTENT INFLAMMATION IN PULMONARY GRANULOMA 48 YEARS AFTER TALCAGE PLEURODESIS, DETECTED BY FDG-PET/CT" BY FANGGIDAY ET AL. STATES THAT FALSE POSITIVE FINDINGS FROM FDG-PET/CT SCANS ARE NOT UNCOMMON. FOREIGN BODY REACTIONS ARE NOT AN UNEXPECTED RESPONSE TO BIOGLUE AND ARE DESCRIBED IN THE PRODUCT'S INSTRUCTIONS FOR USE. RUIZ-ZAFRA ET AL. EXPLAINS THAT "FDG-PET/CT IS HIGHLY SENSITIVE IN THE DETECTION OF NSCLC REOCCURRENCE. HOWEVER, ITS SPECIFICITY IS LIMITED BECAUSE OF THE ELEVATED FDG CONCENTRATIONS CAUSED BY POSTOPERATIVE INFLAMMATION." WHILE BIOGLUE MAY HAVE CAUSED THE INFLAMMATORY RESPONSES IN THE PATIENTS FROM THE PUBLICATION, MISINTERPRETATION OF THE FDG-PET/CT SCANS CAUSED THE UNNECESSARY REOPERATION ON THE PATIENTS. THE AUTHORS MENTION THAT ANOTHER SCAN AT 90 MIN AFTER RADIOTRACER INJECTION CAN BE USED TO MINIMIZE FALSE-POSITIVE RESULTS BECAUSE "INFLAMMATORY CELLS PROGRESSIVELY SUFFER AN FDG WASHOUT, WHICH RESULTS IN AN SUV DECREASE WITH RESPECT TO THE 60-MIN SCAN." METHODS LIKE THE ONE PROPOSED IN THIS PUBLICATION TO RULE OUT FALSE POSITIVE RESULTS SHOULD BE EMPLOYED TO PREVENT UNNECESSARY INVASIVE PROCEDURES FROM BEING PERFORMED.
THIS INVESTIGATION IS CURRENTLY ONGOING. ANY ADDITIONAL INFORMATION WILL BE PROVIDED IN THE FOLLOW-UP REPORT.
ACCORDING TO THE PUBLICATION BY RUIZ-ZAFRA ET AL., THE AUTHORS ALLEGE THAT BIOGLUE CAUSES FALSE POSITIVES IN FDG-PET/CT IMAGING DUE TO FOREIGN BODY REACTIONS TO THE ADHESIVE. FDG-PET/CT IMAGING FOUND A HYPERMETABOLIC PULMONARY NODULE IN THE UPPER LEFT LOBE OF THE PATIENT. LEFT THORACOTOMY AND ATYPICAL PULMONARY RESECTION WERE PERFORMED FOLLOWED BY STAPLER SUTURE. BIOGLUE WAS USED AS AN ADJUNCT TO SUTURE WITH STAPLES. SIX MONTH LATER, FDG-PET/CT IMAGING REVEALED THE PRESENCE OF HYPERMETABOLIC FOCI IN THE RESECTED AREA OF THE LEFT UPPER LOBE, WITH A SUVMAX OF 4.66 SUGGESTIVE OF TUMOR REOCCURRENCE. RE-THORACOTOMY WAS REQUIRED, REVEALING THE PRESENCE OF FIBROTIC TISSUE IN THE SURGICAL SITE. THE SURGEONS FOUND AN ENCAPSULATION SURROUNDING THE AREA WHERE THE BIOADHESIVE WAS APPLIED. HISTOLOGICAL STUDY SHOWED A LESION COMPATIBLE WITH FOREIGN BODY GRANULOMA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11707 | BIOGLUE SURGICAL ADHESIVE | SURGICAL GLUE | MUQ | CRYOLIFE, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Other |