FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 2904774 · Received January 9, 2013

Report

Report Number
2210968-2013-00200
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
December 12, 2012
Report Date
December 13, 2012
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-00201. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE RETURNED BLADE FAILED THE SHARPNESS TEST WITH AN AVERAGE BREAKING FORCE OF 8.03 LBF (SPECIFICATION: AVERAGE 3.5 LBF MAX). THE BLADE WOULD NOT HAVE CUT PROPERLY DURING THE PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, THE RETRACTABLE BLADE DID NOT SPIN WHEN THE TRIGGER WAS SQUEEZED AND THEREFORE COULD NOT CUT. AFTERWARDS, THE BLADE HOUSING STAYED EXTENDED. AT SOME POINT, IT WAS ONLY VIBRATING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11618 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT216473

Patients

Seq Age Sex Outcome Treatment
1