GYNECARE MORCELLEX* TISSUE MORCELLATOR
Report
- Report Number
- 2210968-2013-00200
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 12, 2012
- Report Date
- December 13, 2012
- Manufacturer
- ETHICON, INC.
- Product Code
- HET
- PMA / PMN Number
- K100280
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4): THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. THIS IS ONE OF TWO MEDWATCHES BEING SUBMITTED AS TWO DEVICES WERE INVOLVED IN THIS EVENT. SEE ALSO MEDWATCH 2210968-2013-00201. THE SAME PATIENT IS REPRESENTED IN EACH MEDWATCH.
(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. THE DEVICE WAS VISUALLY AND FUNCTIONALLY EVALUATED. THE RETURNED BLADE FAILED THE SHARPNESS TEST WITH AN AVERAGE BREAKING FORCE OF 8.03 LBF (SPECIFICATION: AVERAGE 3.5 LBF MAX). THE BLADE WOULD NOT HAVE CUT PROPERLY DURING THE PROCEDURE.
IT WAS REPORTED THAT A PATIENT UNDERWENT A LAPAROSCOPIC MYOMECTOMY ON (B)(6) 2012. DURING THE PROCEDURE, THE RETRACTABLE BLADE DID NOT SPIN WHEN THE TRIGGER WAS SQUEEZED AND THEREFORE COULD NOT CUT. AFTERWARDS, THE BLADE HOUSING STAYED EXTENDED. AT SOME POINT, IT WAS ONLY VIBRATING. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 11618 | GYNECARE MORCELLEX* TISSUE MORCELLATOR | LAPAROSCOPE, GYNECOLOGIC | HET | ETHICON, INC. | NA | MT216473 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |