FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2904635 · Received January 9, 2013

Report

Report Number
2531779-2013-00437
Event Type
Malfunction
Date Received
January 9, 2013
Report Date
December 13, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 08/07/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/15/2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE PUMP HISTORY WAS REVIEWED AND SHOWED THE DATES BEGINNING ON 06/08/2013. THE BLACK BOX AND HISTORIES FOR THE COMPLAINT ON 12/14/2012 HAVE BEEN OVERWRITTEN DUE TO CONTINUED USE OF THE PUMP. THERE WERE NO ALARMS ASSOCIATED WITH TO THE COMPLIANT RECORDED IN BLACK BOX OR ALARM HISTORY; ONLY TYPICAL USAGE WAS OBSERVED. THE USER¿S PROGRAMMED BASAL RATES WERE CORRECTLY REFLECTED IN THE DAILY INSULIN DELIVERY TOTALS. THE PUMP PASSED REQUIRED 29 HOUR FLOW ACCURACY TEST SUCCESSFULLY. THE COMPLAINT WAS NOT ABLE TO BE DUPLICATED. THE PUMP INFORMATION FOR THE COMPLAINT DATE HAS BEEN OVERWRITTEN.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED ANIMAS TO REPORT AN ISSUE WITH ALLEGED INACCURATE DELIVERY WITH THE REPORTED INSULIN PUMP. THE TECHNICAL SUPPORT REPRESENTATIVE CONTACTED THE PATIENT TO TROUBLESHOOT THE PUMP; HOWEVER, WAS UNABLE TO REACH HIM BY TELEPHONE. THERE WAS NO PATIENT INJURY ASSOCIATED WITH THIS ISSUE. AS THE ISSUE WAS NOT RESOLVED, THIS COMPLAINT IS BEING REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12215 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION OTP GLUCOSE MGMT SYSTEM

Patients

Seq Age Sex Outcome Treatment
1 68 YR