FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 2904600
·
Received December 5, 2012
Report
- Report Number
- 1811605-2012-00157
- Event Type
- Malfunction
- Date Received
- December 5, 2012
- Date of Event
- November 30, 2012
- Report Date
- December 4, 2012
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
THE USER REPORTED THAT THE PAD STOPPED WORKING. OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO BARE WIRE. CORD BREAKAGE USUALLY OCCURS WHEN THE CORD IS REPEATEDLY OVER BENT NEAR THE SWITCH. WE HAVE INITIATED A CAPA PROJECT ((B)(4)) TO REDUCE THE OCCURRENCE OF THIS ISSUE.
Description of Event or Problem · 1
THE USER REPORTED THAT THE PAD STOPPED WORKING. OUR INSPECTION FOUND A SMALL CUT IN THE CORD NEAR THE SWITCH THAT COULD HAVE EXPOSED THE USER TO BARE WIRE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 055 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |