FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2904585 · Received November 30, 2012

Report

Report Number
3008642652-2012-03183
Event Type
Malfunction
Date Received
November 30, 2012
Date of Event
November 11, 2012
Report Date
November 29, 2012
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION THE MONITOR WAS UNABLE TO POWER UP. THE CAUSE OF THE PROBLEM WAS ISOLATED TO A FRACTURED BGA SOLDER CONNECTION ON RAM CHIPS U100 AND U101. THE ROOT CAUSE OF THE FRACTURED NGA SOLDER CONNECTION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE STRESS ON THE MONITOR C/A BOARD. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCE THE PCA STRAIN WAS SUBMITTED FOR FDA REVIEW ON 11/16/2012 AND IS CURRENTLY UNDER REVIEW. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE COMPONENTS. THE PT RECEIVED A REPLACEMENT MONITOR.

Description of Event or Problem · 1

THE DAUGHTER OF AN (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE PT'S MONITOR WAS MAKING SCREECHING SOUNDS AND WOULDN'T FULLY POWER UP. THE PT WAS ISSUED A REPLACEMENT MONITOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 82 YR