FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2904583
·
Received November 30, 2012
Report
- Report Number
- 3008642652-2012-03181
- Event Type
- Malfunction
- Date Received
- November 30, 2012
- Date of Event
- July 16, 2012
- Report Date
- November 29, 2012
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION OF BATTERY CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (BATTERY CHARGER PROBLEM) HAS BEEN CONFIRMED. UPON INVESTIGATION CONTAMINATION WAS FOUND ON THE CHARGER'S BATTERY BOARD, WHICH PREVENTED THE BATTERY CHARGER/MODEM FROM RECOGNIZING INSERTED BATTERY PACKS. THE ROOT CAUSE FOR THE CONTAMINATION WAS INGRESS OF AN UNKNOWN LIQUID. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED BATTERY BOARD. THE PT RECEIVED A REPLACEMENT CHARGER/MODEM.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS BATTERY CHARGER WAS DISPLAYING "BATTERY CHARGER PROBLEM." THE PT WAS ISSUED A REPLACEMENT BATTERY CHARGER/MODEM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 48 YR |