PROMUS ELEMENT ¿
Report
- Report Number
- 2134265-2012-08292
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- October 26, 2012
- Report Date
- December 12, 2012
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KR
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MFR: A VISUAL AND MICROSCOPIC EXAMINATION FOUND THAT THE STENT AND BALLOON WERE SEVERELY KINKED BETWEEN 12TH AND 13TH PROXIMAL ROWS. SOME STRUTS AT THIS LOCATION WERE BENT OUTWARDS. NO ISSUES WERE NOTED WITH THE TIP OF THE DEVICE THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).
THIS COMPLAINT IS NOW REPORTABLE BASED ON THE ANALYSIS COMPLETED ON (B)(6) 2012. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY INTERVENTION (PCI) THE DEVICE WAS UNABLE TO CROSS THE LESION. THE 80% STENOSED TARGET LESION WAS LOCATED IN THE SEVERELY TORTUOUS AND SEVERELY CALCIFIED LEFT ANTERIOR DESCENDING ARTERY (LAD). THE LESION WAS PREDILATED WITH A 2.0X15 NON-BSC BALLOON CATHETER AND THEN A 2.50X28MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) WAS ADVANCED TO THE LAD; HOWEVER, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE DEVICE WAS REMOVED FROM THE PATIENT AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS STABLE. HOWEVER, RETURNED PRODUCT ANALYSIS REVEALED STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10635 | PROMUS ELEMENT ¿ | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911328250 | 0014415450 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 58 YR |