FDA Adverse Event
Malfunction
Summary report: N
THERMOPHORE AUTOMATIC MOIST HEAT PACK
MDR report key: 2904568
·
Received December 5, 2012
Report
- Report Number
- 1811605-2012-00158
- Event Type
- Malfunction
- Date Received
- December 5, 2012
- Date of Event
- October 1, 2012
- Report Date
- December 4, 2012
- Manufacturer
- BATTLE CREEK EQUIPMENT COMPANY
- Product Code
- IRT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
ANALYSIS OF THE SWITCH COMPONENTS REVEALED THAT SENSITIVITY TO EXTERNAL STRESS HAD CAUSED THE FAILURE OF A RESISTOR. THE FAILURE RESULTED IN A SMALL SPARK THAT WAS CONTAINED BY OUR DOUBLE-INSULATED DESIGN. ALTHOUGH THIS IS NOT A STATISTICALLY SIGNIFICANT COMPLAINT AND MAY BE DUE TO MISUSE OR FAILURE TO FOLLOW INSTRUCTIONS OF THE PART OF THE CONSUMER, WE HAVE INITIATED THE FOLLOWING CORRECTIVE ACTIONS: WE HAVE REINFORCED THE TERMINALS LEADING INTO THE SWITCH TO REDUCE THE POSSIBILITY OF CURRENT FLUCTUATIONS AND/OR OPEN CONTINUITY. WE HAVE REPOSITIONED COMPONENTS TO REDUCE THE POSSIBILITY OF ACCIDENTAL CONTACT BETWEEN LIVE COMPONENTS.
Description of Event or Problem · 1
THE USER REPORTED THAT THE UNIT CAUSED A SPARK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | THERMOPHORE AUTOMATIC MOIST HEAT PACK | ELECTRIC MOIST HEAT PACK | IRT | BATTLE CREEK EQUIPMENT COMPANY | 095 |