FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 2904507 · Received January 9, 2013

Report

Report Number
3005075853-2013-00156
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
January 2, 2013
Report Date
January 3, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME. ADDITIONAL INFORMATION REQUESTED: WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? ASKU. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? ASKU. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? UNKNOWN. WAS THERE ANY TORQUE OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? UNKNOWN. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? UNKNOWN. WERE ANY UNEXPECTED NOISES HEARD? UNKNOWN. DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? UNKNOWN. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? UNKNOWN.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RECEIVED IN GOOD VISUAL CONDITION. UPON CYCLING THE DEVICE, IT WAS NOTED TO BE EMPTY AND LOCKED OUT. THE INSTRUMENT IS DESIGNED TO LOCK OUT AFTER ALL THE CLIPS HAVE BEEN FIRED; THEREFORE, A POTENTIAL CAUSE OF THE CUSTOMER REPORTED EXPERIENCE IS THE FIRING OF ALL OF THE CLIPS AND THE INSTRUMENT "WILL NOT FIRE" (ACTIVATION OF THE LOCK OUT MECHANISM). THE INSTRUMENT HAS AN ORANGE INDICATOR THAT APPEARS ON THE TOP OF THE HANDLE AS A REFERENCE FOR THE USER AS TO THE QUANTITY OF CLIPS REMAINING. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE EVENT REPORTED. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE HAD FIRING ISSUES. ONE CLIP SCISSORED AND ANOTHER WAS NOT CLIPPING ONTO THE TISSUE. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO ADVERSE CONSEQUENCES REPORTED. NO OTHER DETAILS ARE AVAILABLE.