FDA Adverse Event Injury Summary report: N

PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM

MDR report key: 2904233 · Received January 9, 2013

Report

Report Number
2024168-2013-00163
Event Type
Injury
Date Received
January 9, 2013
Date of Event
November 29, 2012
Report Date
December 17, 2012
Manufacturer
AV-TEMECULA-CT
Product Code
MGB
PMA / PMN Number
P960043
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SZ
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. IT WAS REPORTED THAT THE PROGLIDE DEVICE WAS USED IN A CALCIFIED VESSEL. PER THE INSTRUCTIONS FOR USE, THE SAFETY AND EFFECTIVENESS OF THE PROGLIDE DEVICE HAVE NOT BEEN ESTABLISHED FOR PATIENTS WITH FEMORAL ARTERY CALCIUM WHICH IS FLUOROSCOPICALLY VISIBLE AT THE ACCESS SITE. THE CUSTOMER REPORTED THE DEVICE WAS DISCARDED. A FOLLOW UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.

Additional Manufacturer Narrative · 1

(B)(4). PER THE INSTRUCTIONS FOR USE (IFU), THIS DEVICE SHOULD ONLY BE USED BY PHYSICIANS (OR OTHER HEALTHCARE PROFESSIONALS AUTHORIZED BY OR UNDER THE DIRECTION OF SUCH PHYSICIANS) WHO ARE TRAINED IN DIAGNOSTIC AND THERAPEUTIC CATHETERIZATION PROCEDURES AND WHO HAVE BEEN TRAINED BY AN AUTHORIZED REPRESENTATIVE OF ABBOTT VASCULAR.

Additional Manufacturer Narrative · 1

(B)(4). ESTIMATED DATE OF OCCURRENCE, BASED ON REPORT OF INCIDENT OCCURRING WITHIN 4 WEEKS PRIOR TO (B)(6) 2013.

Additional Manufacturer Narrative · 1

(B)(4). IT IS INDICATED THAT THE DEVICE IS NOT RETURNING FOR EVALUATION; THEREFORE, A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES THAT WOULD HAVE CONTRIBUTED TO THE REPORTED EVENT. RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FROM THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ARTERIOTOMY CLOSURE OF A MODERATELY CALCIFIED COMMON FEMORAL ARTERY WAS ATTEMPTED USING A PROGLIDE DEVICE AFTER AN INTERVENTIONAL PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY PROCEDURE. REPORTEDLY, WHEN TRYING TO PULL BACK THE WIRE OVER THE SYSTEM, THE WHOLE SYSTEM WAS STUCK ON THE WIRE AND HAD TO BE REMOVED. HEMOSTASIS WAS ACHIEVED BY APPLYING MANUAL ARTERIAL COMPRESSION. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA OR CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. THE PHYSICIAN IS REPORTEDLY TRAINED IN THE USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 1

SUBSEQUENT TO THE INITIAL MEDWATCH REPORT, ADDITIONAL INFORMATION RECEIVED INDICATED THAT THE PHYSICIAN IS NOT TRAINED IN USE OF THE PROGLIDE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12855 PERCLOSE PROGLIDE 6F SUTURE MEDIATED CLOSURE (SMC) SYSTEM SUTURE MEDIATED CLOSURE MGB AV-TEMECULA-CT 21006J1

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention SHEATH: 8F| SHEATH: 8F