FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 2904159 · Received January 3, 2013

Report

Report Number
2518422-2013-00001
Event Type
Malfunction
Date Received
January 3, 2013
Date of Event
December 19, 2012
Report Date
December 19, 2012
Manufacturer
RESPIRONICS INC
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SVC CTR, A FAILURE OF THE DEVICE TO POWER ON WAS OBSERVED. THE DEVICE'S SYS BOARD WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

THE MFR RECEIVED INFO ALLEGING A VENTILATOR WOULD NOT POWER ON. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
3471 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS INC 1054260

Patients

Seq Age Sex Outcome Treatment
1