TENACULUM FORCEPS
Report
- Report Number
- 2955842-2013-00101
- Event Type
- Malfunction
- Date Received
- January 9, 2013
- Date of Event
- December 10, 2012
- Report Date
- February 22, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING FOUND THE INSTRUMENT HAD A BROKEN CABLE AT THE DISTAL END. THE INSTRUMENT WAS FOUND WITH A FRAYED PITCH CABLE AT THE DISTAL CLEVIS HUB. NO DAMAGE WAS FOUND AT THE CLEVIS. THE FRAYED SEGMENT WAS APPROXIMATELY 0.02 IN LENGTH. NO OTHER CABLE DAMAGE WAS FOUND.
THE INSTRUMENT AND/OR ACCESSORY HAVE NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE ROOT CAUSE OF THE CUSTOMER REPORTED FAILURE MODE CANNOT BE DETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF THE INSTRUMENT IS RETURNED (POST ENGINEERING EVALUATION), OR IF ADDITIONAL INFORMATION IS RECEIVED.
IT WAS REPORTED THAT AFTER A DA VINCI SI CYSTECTOMY PROCEDURE, A BROKEN CABLE AT THE DISTAL END OF THE TENACULUM FORCEPS INSTRUMENT WAS DISCOVERED. THE INTENDED PROCEDURE WAS COMPLETED SUCCESSFULLY. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 10919 | TENACULUM FORCEPS | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420207-06 | M10120528 403 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |