SYMBIQ SINGLE CHANNE
Report
- Report Number
- 9615050-2013-00005
- Event Type
- Malfunction
- Date Received
- January 3, 2013
- Date of Event
- December 4, 2012
- Report Date
- December 4, 2012
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K110901
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- HI, US
- Reporter Occupation
- OTHER
Narratives
TESTING AND INVESTIGATION FOUND AT POWER UP, THE DEVICE DISPLAYED A WHITESCREEN ERROR AND THE DEVICE SOUNDED AN AUDIBLE ALARM FROM THE SECONDARY BEEPER. THIS WAS DUE TO A SOFTWARE ERROR. THIS REPORT REPRESENTS ALL THE INFO KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED A WHITESCREEN ERROR. THE PUMP WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED "SOFTWARE ERROR." NO TRACKING INFO WAS PROVIDED; THEREFORE, SPECIFIC PT INFO, PUMP PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PT EVENTS AND NO REPORTED DELAYS OF CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, AFTER THE DEVICE WA POWERED ON, THE DEVICE ALARMED WITH A WHITESCREEN ERROR. THOUGH REQUESTED, NO ADD'L INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 3384 | SYMBIQ SINGLE CHANNE | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |