FDA Adverse Event
Malfunction
Summary report: N
SECURE II MED/SURG BED
MDR report key: 2903975
·
Received November 30, 2012
Report
- Report Number
- 1831750-2012-12415
- Event Type
- Malfunction
- Date Received
- November 30, 2012
- Date of Event
- November 3, 2012
- Report Date
- November 3, 2012
- Manufacturer
- STRYKER MEDICAL
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
SIDERAIL GLIDE RAILS DIRTY, GLIDE ROD GLIDE BUSHINGS PUSHED OUT.
Description of Event or Problem · 1
IT WAS REPORTED BY SERVICE REPORT THAT THE SIDE RAIL WOULD NOT LOCK IN THE UPRIGHT POSITION. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES ARE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SECURE II MED/SURG BED | A/C POWERED HOSPITAL BED | FNL | STRYKER MEDICAL | 3002 EX | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|