FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2903700 · Received January 9, 2013

Report

Report Number
1416980-2013-00597
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
December 13, 2012
Report Date
December 17, 2012
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FKX
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS IS A PRODUCT NOT DISTRIBUTED IN THE US AND DOES NOT HAVE A 510K NUMBER. A REQUEST FOR THE RETURN OF THE DEVICE HAS BEEN MADE. SHOULD THE DEVICE BE RECEIVED BY BAXTER FOR EVALUATION, A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF AN EVALUATION OR IF ANY ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE SAMPLE WAS RECEIVED AND EVALUATED. A VISUAL INSPECTION WAS PERFORMED AND A PINHOLE WAS FOUND IN THE CASSETTE. DURING PRESSURE TESTING A LEAK WAS NOTED TO BE COMING FROM THE PINHOLE. THIS COMPLAINT FOR A REPORT OF A LEAK WAS CONFIRMED. THE ROOT CAUSE WAS UNDETERMINED. A FOLLOW-UP MDR WILL BE SUBMITTED IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

ON (B)(6) 2012, A DOCTOR REPORTED THAT ON (B)(6) 2012 A LEAK WAS NOTED COMING FROM A CASSETTE DURING USE. THE ACTUAL SAMPLE WAS AVAILABLE AND REQUESTED FOR EVALUATION. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
12162 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE