FDA Adverse Event Malfunction Summary report: N

TAXUS¿ LIBERTÉ¿

MDR report key: 2903578 · Received January 9, 2013

Report

Report Number
2134265-2012-08156
Event Type
Malfunction
Date Received
January 9, 2013
Date of Event
September 25, 2012
Report Date
December 11, 2012
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
EG
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MFR.: VISUAL EXAMINATION OF THE RETURNED DEVICE NOTED DISTAL STENT STRUT DAMAGE AND A NUMBER OF STRUTS WERE BENT BACK PROXIMALLY. THE BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. FURTHER EXAMINATION ALSO NOTED SLIGHT KINKING OF THE MID SHAFT REGION. NO FURTHER ISSUES WERE NOTED WITH THE RETURNED DEVICE WHICH COULD HAVE CONTRIBUTED TO THE REPORTED COMPLAINT. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2012. IT WAS REPORTED THAT DURING A STENTING TREATMENT PROCEDURE A STENT WAS UNABLE TO CROSS THE LESION. THE 2.5 X 16MM TAXUS LIBERTE STENT DELIVERY SYSTEM WAS ADVANCED MANY TIMES TO AN UNSPECIFIED LESION. DESPITE A GUIDE CATHETER EXCHANGE, AND THE BUDDY WIRE TECHNIQUE, THE DEVICE WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. HOWEVER, DEVICE ANALYSIS REVEALED STENT DAMAGE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11311 TAXUS¿ LIBERTÉ¿ CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - GALWAY H7493894016250 14764962

Patients

Seq Age Sex Outcome Treatment
1