FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2903365 · Received December 14, 2012

Report

Report Number
1720753-2012-10473
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
December 5, 2012
Report Date
December 14, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE CONTROL PANEL PROCESSOR CIRCUIT BOARD WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A COMMUNICATIONS FAIL AND CONTROL PANEL PROCESSOR ERROR MESSAGE. THESE MESSAGES WILL LIKELY RESULT IN A SYSTEM SHUT DOWN SITUATION AND/OR LOCK UP DEPENDING ON WHEN THE ERROR MESSAGE OCCURRED. THERE IS NO REPORT OF PATIENT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1