FDA Adverse Event Malfunction Summary report: N

KNIFE

MDR report key: 2903332 · Received December 14, 2012

Report

Report Number
2523835-2012-00137
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
November 16, 2012
Report Date
November 16, 2012
Manufacturer
ALCON PRECISION DEVICE - SINKING SPRING
Product Code
HNN
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MY
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION IS IN PROGRESS. SAMPLES HAVE NOT YET BEEN RECEIVED FOR EVALUATION. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MANUFACTURER'S ACCEPTANCE CRITERIA. A ROOT CAUSE HAS NOT BEEN IDENTIFIED. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED THAT THE KNIFE WAS NOT SHARP AND THE SHAPE WAS "NOT GOOD" DURING SURGERY. THE KNIFE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO HARM TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 KNIFE MANUAL OPHTHALMIC SURGICAL INSTRUMENT HNN ALCON PRECISION DEVICE - SINKING SPRING 8065982265 803686M

Patients

Seq Age Sex Outcome Treatment
1