FDA Adverse Event
Malfunction
Summary report: N
KNIFE
MDR report key: 2903332
·
Received December 14, 2012
Report
- Report Number
- 2523835-2012-00137
- Event Type
- Malfunction
- Date Received
- December 14, 2012
- Date of Event
- November 16, 2012
- Report Date
- November 16, 2012
- Manufacturer
- ALCON PRECISION DEVICE - SINKING SPRING
- Product Code
- HNN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MY
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
THE INVESTIGATION IS IN PROGRESS. SAMPLES HAVE NOT YET BEEN RECEIVED FOR EVALUATION. THE DEVICE HISTORY RECORD (DHR) FOR THE LOT WAS REVIEWED. FUNCTIONAL PENETRATION TEST VALUES FOR THE LOT MET SPECIFICATION. NO ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO A DULL KNIFE WERE FOUND DURING THE DHR REVIEW AND THE PRODUCT WAS RELEASED ACCORDING TO THE MANUFACTURER'S ACCEPTANCE CRITERIA. A ROOT CAUSE HAS NOT BEEN IDENTIFIED. (B)(4).
Description of Event or Problem · 1
A CUSTOMER REPORTED THAT THE KNIFE WAS NOT SHARP AND THE SHAPE WAS "NOT GOOD" DURING SURGERY. THE KNIFE WAS EXCHANGED AND THE PROCEDURE WAS COMPLETED WITH NO HARM TO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | KNIFE | MANUAL OPHTHALMIC SURGICAL INSTRUMENT | HNN | ALCON PRECISION DEVICE - SINKING SPRING | 8065982265 | 803686M |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |