FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2903329 · Received December 14, 2012

Report

Report Number
1720753-2012-10486
Event Type
Malfunction
Date Received
December 14, 2012
Date of Event
September 4, 2012
Report Date
December 14, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY AND THE LEMO PIN CONNECTOR WERE REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED AN INTERMITTENT COMMUNICATION ERRORS, WHICH PREVENTED THE SYSTEM FROM BOOTING UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.