FDA Adverse Event Injury Summary report: N

REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7

MDR report key: 2903298 · Received January 9, 2013

Report

Report Number
0002249697-2013-00009
Event Type
Injury
Date Received
January 9, 2013
Date of Event
December 19, 2012
Report Date
December 19, 2012
Manufacturer
STRYKER ORTHOPAEDICS-MAHWAH
Product Code
MEH
PMA / PMN Number
K071082
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION HAS BEEN REQUESTED AND IF RECEIVED WILL BE PROVIDED IN A SUPPLEMENTAL REPORT.

Additional Manufacturer Narrative · 1

AN EVENT REGARDING ALTR INVOLVING A REJUVENATE MODULAR DEVICE WAS REPORTED. THE EVENT WAS CONFIRMED. DEVICE EVALUATION WAS NOT PERFORMED AS NO DEVICES WERE RECEIVED. A DEVICE HISTORY REVIEW FOUND THAT THE DEVICES IN THE REPORTED LOT WERE ACCEPTED INTO FINAL STOCK WITH NO REPORTED DISCREPANCIES. THE SUPER AND CHS COMPLAINT DATABASES SHOW THERE HAVE BEEN OTHER EVENTS FOR THE REPORTED LOT. SIMILAR EVENTS HAVE OCCURRED FOR THE CATALOG NUMBER AND PRODUCT FAMILY. THESE EVENTS WERE DETERMINED TO BE ASSOCIATED WITH RA 2012-067. VOLUNTARY RECALL RA 2012-067 WAS INITIATED FOR ABGII AND REJUVENATE MODULAR STEMS AND NECKS DUE TO THE POTENTIAL RISKS ASSOCIATED WITH THESE DEVICES. THE REPORTED ALTR IS CONSIDERED TO BE UNDER THE SCOPE OF THIS RECALL. NO FURTHER INVESTIGATION IS REQUIRED.

Description of Event or Problem · 1

PATIENT COMPLAINED OF PAIN AND HAD ELEVATED COBALT SO THE SURGEON REVISED.

Description of Event or Problem · 1

PATIENT COMPLAINED OF PAIN AND HAD ELEVATED COBALT SO THE SURGEON REVISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
11814 REJUVENATE STRGHT PRFIT TMZF MOD STEM SIZE 7 IMPLANT MEH STRYKER ORTHOPAEDICS-MAHWAH NA MKA2T8

Patients

Seq Age Sex Outcome Treatment
1 64 YR Required Intervention