CHARGING SYSTEM
Report
- Report Number
- 1627487-2013-13001
- Event Type
- Injury
- Date Received
- January 2, 2013
- Date of Event
- December 4, 2012
- Report Date
- December 4, 2012
- Manufacturer
- ST JUDE MED - NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-001-C
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVAL: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPT TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPG'S AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MED HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MED HISTORY.
DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-13002. IT WAS REPORTED BY THE PT THE CHARGING SYS WOULD GET HOT WHILE CHARGING. MULTIPLE ATTEMPTS TO CONTACT THE PT WERE UNSUCCESSFUL.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 306 | CHARGING SYSTEM | SCS CHARGING SYSTEM | LGW | ST JUDE MED - NEUROMODULATION | 3721 | 2791325 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR | Required Intervention | SCS LEAD, MODEL 3146 (2)| IMPLANT:| SCS EXTENSION, MODEL 3342| IMPLANT:| IMPLANT:| SCS LEAD, MODEL 3186 |