FDA Adverse Event Injury Summary report: N

CHARGING SYSTEM

MDR report key: 2903194 · Received January 2, 2013

Report

Report Number
1627487-2013-13001
Event Type
Injury
Date Received
January 2, 2013
Date of Event
December 4, 2012
Report Date
December 4, 2012
Manufacturer
ST JUDE MED - NEUROMODULATION
Product Code
LGW
PMA / PMN Number
P010032
Removal / Correction Number
1627487-07262012-001-C
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THIS CHARGER MODEL WAS ASSOCIATED WITH A FIELD CORRECTION. MFR'S EVAL: CORRECTIVE AND PREVENTIVE ACTION (CAPA) INVESTIGATION WAS PERFORMED. EVAL: RESULTS: POCKET HEATING WAS CONFIRMED. THE INVESTIGATION FOR CAPA (B)(4) ASSOCIATED WITH HEATING WHILE CHARGING (POCKET HEATING) CONCLUDED THAT THE CHARGER WAS CAPABLE OF TRANSFERRING ENERGY TO THE IPG AT A RATE THAT WOULD CAUSE HEATING OF THE IPG AND/OR CHARGING WAND OF SUFFICIENT ELEVATED TEMPT TO CAUSE PAIN AND BURNS. THE HEATING WHILE CHARGING WAS DETERMINED TO BE EXACERBATED BY OFF-AXIS CHARGING OF SHALLOW IMPLANTED IPG'S AND THAT ALL CHARGERS WERE CAPABLE OF ELEVATED HEATING. SJM HAS LIMITED INFO RELATED TO THE PT'S MED HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PT'S PHYSICIAN REGARDING MED HISTORY.

Description of Event or Problem · 1

DEVICE 1 OF 2. REFERENCE MFR REPORT: 1627487-2013-13002. IT WAS REPORTED BY THE PT THE CHARGING SYS WOULD GET HOT WHILE CHARGING. MULTIPLE ATTEMPTS TO CONTACT THE PT WERE UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
306 CHARGING SYSTEM SCS CHARGING SYSTEM LGW ST JUDE MED - NEUROMODULATION 3721 2791325

Patients

Seq Age Sex Outcome Treatment
1 47 YR Required Intervention SCS LEAD, MODEL 3146 (2)| IMPLANT:| SCS EXTENSION, MODEL 3342| IMPLANT:| IMPLANT:| SCS LEAD, MODEL 3186