FDA Adverse Event Malfunction Summary report: N

TPS UNIVERSAL DRIVER

MDR report key: 2903154 · Received November 26, 2012

Report

Report Number
1811755-2012-04167
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
October 29, 2012
Report Date
November 14, 2012
Manufacturer
STRYKER INSTRUMENTS, INSTRUMENTS DIV.
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

SEVERAL THIRD PARTY PARTS WERE FOUND DURING DEVICE EVAL, INCLUDING THE MOTOR ASSEMBLY AND THE POTTED TRIGGER, WHICH ARE PROBABLE CAUSES OF THE DEVICE RUNNING WITHOUT USER ACTIVATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MFR FACILITY, THE DEVICE WAS RUNNING AT LOW SPEED WITHOUT USER ACTIVATION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TPS UNIVERSAL DRIVER ERL STRYKER INSTRUMENTS, INSTRUMENTS DIV.

Patients

Seq Age Sex Outcome Treatment
1 UNK