FDA Adverse Event
Malfunction
Summary report: N
TPS UNIVERSAL DRIVER
MDR report key: 2903154
·
Received November 26, 2012
Report
- Report Number
- 1811755-2012-04167
- Event Type
- Malfunction
- Date Received
- November 26, 2012
- Date of Event
- October 29, 2012
- Report Date
- November 14, 2012
- Manufacturer
- STRYKER INSTRUMENTS, INSTRUMENTS DIV.
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
SEVERAL THIRD PARTY PARTS WERE FOUND DURING DEVICE EVAL, INCLUDING THE MOTOR ASSEMBLY AND THE POTTED TRIGGER, WHICH ARE PROBABLE CAUSES OF THE DEVICE RUNNING WITHOUT USER ACTIVATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING TESTING CONDUCTED AT THE MFR FACILITY, THE DEVICE WAS RUNNING AT LOW SPEED WITHOUT USER ACTIVATION. NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE ALLEGED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | TPS UNIVERSAL DRIVER | ERL | STRYKER INSTRUMENTS, INSTRUMENTS DIV. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |