FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 2902825 · Received January 9, 2013

Report

Report Number
3008382007-2013-00375
Event Type
Malfunction
Date Received
January 9, 2013
Report Date
December 17, 2012
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 ((B)(4) 2013)-DEVICE EVALUATION: THE PRODUCTS INVOLVED WITH THIS COMPLAINT HAVE BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS (PA) RESPECTIVELY ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE TEST STRIPS INVOLVED WITH THIS COMPLAINT WERE EVALUATED AND ER4 WAS OBSERVED WITH CONTROL SOLUTION TESTING. THE REPORTED COMPLAINT WAS OBSERVED ON THE METER'S ERROR LOG; HOWEVER, PA WAS UNABLE TO REPRODUCE FAILURE IN METER TESTING. IF LIFESCAN OBTAINS ADDITIONAL INFORMATION REGARDING THIS COMPLAINT, A FOLLOW UP REPORT WILL BE SUBMITTED. AT THIS TIME, LIFESCAN CONSIDERS THIS MATTER CLOSED.

Additional Manufacturer Narrative · 1

THE SUBJECT METER AND TEST STRIPS HAVE BEEN RETURNED TO LIFESCAN FOR EVALUATION. THE EVALUATION OF THE PRODUCTS HAVE NOT BEEN COMPLETED; THEREFORE, NO CONCLUSIONS CAN BE DRAWN AT THIS TIME. ONCE THE EVALUATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILLED.

Description of Event or Problem · 1

ON (B)(6) 2012, THE REPORTER CONTACTED LIFESCAN USA, ALLEGING ERROR 4. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
10500 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3303961

Patients

Seq Age Sex Outcome Treatment
1