FDA Adverse Event Malfunction Summary report: N

RIATA ACTIVE FIXATION

MDR report key: 2902650 · Received January 8, 2013

Report

Report Number
2017865-2013-00290
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
November 14, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
Z0457
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Additional Manufacturer Narrative · 1

A PARTIAL LEAD WAS RETURNED FOR ANALYSIS. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE FOUND AT 7.5-11.2CM FROM THE DISTAL TIP. THE SILICONE INSULATION WAS ABRADED AND THE SENSING CONDUCTOR WAS VISIBLE. THE ETFE COATING WAS INTACT AT THE SAME LOCATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN THE OPERATING ROOM FOR DEVICE CHANGE-OUT DUE TO NORMAL ERI. EXTERNALIZED CONDUCTORS WERE NOTED VIA FLUOROSCOPY. LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9273 RIATA ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 1580/60 NA

Patients

Seq Age Sex Outcome Treatment
1 44 YR