FDA Adverse Event
Malfunction
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR
MDR report key: 2902639
·
Received January 8, 2013
Report
- Report Number
- 2017865-2013-00385
- Event Type
- Malfunction
- Date Received
- January 8, 2013
- Date of Event
- October 1, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING DEVICE CHANGE OUT, LOW SENSING WAS OBSERVED. LEAD DISLODGEMENT WAS SUSPECTED. WHEN REPOSITIONING ATTEMPTS RESULTED IN POOR SENSING AND IMPEDANCE VALUES, THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 8507 | DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7120Q/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |