FDA Adverse Event
Injury
Summary report: N
DURATA STS OPTIM ACTIVE FIXATION
MDR report key: 2902564
·
Received January 8, 2013
Report
- Report Number
- 2017865-2013-00395
- Event Type
- Injury
- Date Received
- January 8, 2013
- Date of Event
- October 29, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
NO MEDWATCH FORM WAS RECEIVED.
Additional Manufacturer Narrative · 1
ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND. A LEAD TIP STIFFNESS TEST WAS PERFORMED AND AND THE LEAD WAS FOUND TO BE WITHIN SPECIFICATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT POST IMPLANT, THE PATIENT EXPERIENCED CHEST PAIN. A LEAD PERFORATION WITH PERICARDIAL EFFUSION WAS OBSERVED VIA ECHO. WHEN REPOSITIONING WAS UNSUCCESSFUL, THE LEAD WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 8757 | DURATA STS OPTIM ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7121/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | Required Intervention |