FDA Adverse Event Injury Summary report: N

DURATA STS OPTIM ACTIVE FIXATION

MDR report key: 2902564 · Received January 8, 2013

Report

Report Number
2017865-2013-00395
Event Type
Injury
Date Received
January 8, 2013
Date of Event
October 29, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
UT
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

NO MEDWATCH FORM WAS RECEIVED.

Additional Manufacturer Narrative · 1

ANALYSIS WAS NORMAL. NO ANOMALY WAS FOUND. A LEAD TIP STIFFNESS TEST WAS PERFORMED AND AND THE LEAD WAS FOUND TO BE WITHIN SPECIFICATION.

Description of Event or Problem · 1

IT WAS REPORTED THAT POST IMPLANT, THE PATIENT EXPERIENCED CHEST PAIN. A LEAD PERFORATION WITH PERICARDIAL EFFUSION WAS OBSERVED VIA ECHO. WHEN REPOSITIONING WAS UNSUCCESSFUL, THE LEAD WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8757 DURATA STS OPTIM ACTIVE FIXATION DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7121/60 NA

Patients

Seq Age Sex Outcome Treatment
1 76 YR Required Intervention