Narratives

Additional Manufacturer Narrative · 1

INTERNAL INSULATION ABRASION WAS NOTED AT 9.6-10.2CM FROM THE ELECTRODE TIP. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 11.1-13.9CM FROM THE ELECTRODE TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING DEVICE CHANGE OUT, EXTERNALIZED CONDUCTORS WERE OBSERVED VIA CINE. THE LEAD WAS EXPLANTED.