FDA Adverse Event
Malfunction
Summary report: N
RIATA ST ACTIVE FIXATION
MDR report key: 2902560
·
Received January 8, 2013
Report
- Report Number
- 2017865-2013-00369
- Event Type
- Malfunction
- Date Received
- January 8, 2013
- Date of Event
- November 8, 2012
- Manufacturer
- ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
- Product Code
- NVY
- PMA / PMN Number
- P950022
- Removal / Correction Number
- Z0458
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
INTERNAL INSULATION ABRASION WAS NOTED AT 9.6-10.2CM FROM THE ELECTRODE TIP. EXTERNALIZED CONDUCTORS DUE TO INTERNAL INSULATION ABRASION WERE NOTED AT 11.1-13.9CM FROM THE ELECTRODE TIP. THE ETFE COATING WAS INTACT AT THESE LOCATIONS.
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. DEVICE EVALUATION ANTICIPATED BUT NOT YET BEGUN.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING DEVICE CHANGE OUT, EXTERNALIZED CONDUCTORS WERE OBSERVED VIA CINE. THE LEAD WAS EXPLANTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 9808 | RIATA ST ACTIVE FIXATION | DEFIBRILLATION LEAD | NVY | ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION | 7002/60 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR |