FDA Adverse Event Malfunction Summary report: N

DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR

MDR report key: 2902521 · Received January 8, 2013

Report

Report Number
2017865-2013-00389
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
October 23, 2012
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVY
PMA / PMN Number
P950022
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. THE DAMAGE FOUND WAS SUSTAINED DURING THE SURGICAL PROCEDURE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE ER AFTER RECEIVING SEVERAL INAPPROPRIATE SHOCKS. UPON INTERROGATION OF THE DEVICE, A LOSS OF CAPTURE AND HIGH PACING THRESHOLD WERE NOTED. X-RAY REVEALED PERFORATIONAND DISLODGEMENT. LEAD WAS EXPLANTED WHEN REPOSITIONING WAS UNSUCCESSFUL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8716 DURATA STS OPTIM ACTIVE FIXATION, DF-4 CONNECTOR DEFIBRILLATION LEAD NVY ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION 7120Q/58 NA

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention