Summary report: N
LARGE NEEDLE DRIVER INSTRUMENT
MDR report key: 2902495
·
Received January 8, 2013
Report
- Report Number
- 2955842-2013-00096
- Event Type
- Malfunction
- Date Received
- January 8, 2013
- Date of Event
- December 11, 2012
- Report Date
- December 12, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT BOTH OF THE PITCH CABLES ARE FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STICK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING OF THE LARGE NEEDLE DRIVER INSTRUMENT, IT WAS OBSERVED THAT THE WIRE ON THE INSTRUMENT WAS BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 9993 | LARGE NEEDLE DRIVER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420006-06 | M10120329 122 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES. |