FDA Adverse Event Malfunction Summary report: N

LARGE NEEDLE DRIVER INSTRUMENT

MDR report key: 2902495 · Received January 8, 2013

Report

Report Number
2955842-2013-00096
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
December 11, 2012
Report Date
December 12, 2012
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, ENGINEERING OBSERVED THAT BOTH OF THE PITCH CABLES ARE FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STICK OUT AT THE WRIST. THE OTHER CABLES AT THE WRIST ARE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING REPROCESSING OF THE LARGE NEEDLE DRIVER INSTRUMENT, IT WAS OBSERVED THAT THE WIRE ON THE INSTRUMENT WAS BROKEN. THERE WERE NO MISSING OR FALLEN PIECES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9993 LARGE NEEDLE DRIVER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420006-06 M10120329 122

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SURGICAL SYSTEM AND ACCESSORIES.