FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2902441 · Received January 8, 2013

Report

Report Number
3004209178-2013-90120
Event Type
Injury
Date Received
January 8, 2013
Date of Event
December 23, 2012
Report Date
December 24, 2012
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
OR
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT TEST. THE DEVICE HAD SCRATCHED DISPLAY WINDOW, BROKEN BELT CLIP SLOT, AND CRACKED RESERVOIR TUBE. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED BEING IN THE EMERGENCY ROOM DUE TO HIGH BLOOD GLUCOSE OF 487MG/DL. THE CUSTOMER EXPERIENCED TROUBLE BREATHING, NAUSEA, AND DEHYDRATION. TROUBLESHOOTING WAS PERFORMED. ASSISTED THE CALLER TO RUN A HIGH PRESSURE TEST AND THE TEST PASSED. THE CUSTOMER MENTIONED THAT SHE WAS IN A CAR ACCIDENT TWO DAYS BEFORE HER ADMISSION, AND SHE WAS THE DRIVER. THE CUSTOMER STATED THAT SHE WAS DISCONNECTED FROM THE INSULIN PUMP PER DOCTOR'S INSTRUCTIONS, AND THE DEVICE WAS DAMAGED AT THE RESERVOIR COMPARTMENT DURING THE ACCIDENT. ADVISED THE CUSTOMER TO DISCONTINUE USE OF THE DEVICE AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9327 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAS

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization