FDA Adverse Event
Malfunction
Summary report: N
FENESTRATED BIPOLAR FORCEPS INSTRUMENT
MDR report key: 2902356
·
Received January 8, 2013
Report
- Report Number
- 2955842-2013-00091
- Event Type
- Malfunction
- Date Received
- January 8, 2013
- Date of Event
- December 14, 2012
- Report Date
- December 14, 2012
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NE
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. ENGINEERING OBSERVED THE PITCH CABLE IS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP STILL REMAINED IN THE CLEVIS. THE CLEVIS DOES NOT EXHIBIT EXCESSIVE DAMAGE. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING REPROCESSING THE FENESTRATED BIPOLAR FORCEPS INSTRUMENT WERE NOTED TO HAVE A FRAYED CABLE. NOTHING WAS REPORTED HAVING FALLEN INTO THE PATIENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 8996 | FENESTRATED BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420205-05 | M10121012 325 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES |