FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2902139 · Received January 8, 2013

Report

Report Number
1416980-2013-00527
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
December 16, 2012
Report Date
December 16, 2012
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PROBLEM WAS CONFIRMED. THE ROOT CAUSE WAS UNDETERMINED. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USE ERROR IN THIS COMPLAINT. THERE WAS NO ALLEGATION REPORTED AGAINST THE BAXTER PRODUCT BY THE CUSTOMER; THEREFORE, THE SAMPLE WAS NOT REQUESTED FOR EVALUATION AND A SAMPLE EVALUATION WILL NOT BE CONDUCTED. THE LOT NUMBER WAS NOT PROVIDED, THEREFORE NO BATCH REVIEW COULD BE PERFORMED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED A BAXTER TECHNICAL SERVICE REPRESENTATIVE (TSR) REGARDING A CHECK HEATER LINE ALARM THAT APPEARED ON THE DISPLAY OF THE HOME PATIENT'S (HP) HOMECHOICE (HC) MACHINE DURING FILL 1, DURING USE, PATIENT CONNECTED. FILL VOLUME = 61 ML. THE HP STATED THAT SHE HAD CHANGED OUT JUST THE HEATER BAG PER THE REGISTERED NURSES ADVICE. THE TSR EXPLAINED TO THE HP THAT THE SETUP IS COMPROMISED, ASSISTED THE HP IN ENDING THE THERAPY AND ADVISED THE HP TO START OVER WITH NEW SUPPLIES. THE HC IS OPERATIONAL. PER INITIAL REPORT, BAXTER'S TECHNICAL SERVICE CENTER ASSISTED THE CUSTOMER IN EVALUATING AND RESOLVING THE ALARM DURING THE INITIAL CONTACT. THERE WAS PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT. PRODUCT SURVEILLANCE RECEIVED A CALL FORM THE HOME PATIENT ON (B)(6) 2013 IN AND SHE SAID THAT SHE HAS BEEN COMPLETING THERAPY SUCCESSFULLY SINCE THEN. SHE STARTED OVER WITH NEW SUPPLIES THAT DAY AFTER CONTACTING BAXTER. THERE WAS PATIENT INVOLVEMENT BUT NO REPORTED INJURY OR MEDICAL INTERVENTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8840 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 74 YR HOME CHOICE