FDA Adverse Event Malfunction Summary report: N

CORE UNIVERSAL DRIVER

MDR report key: 2902032 · Received November 26, 2012

Report

Report Number
1811755-2012-04151
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
October 24, 2012
Report Date
October 24, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY, IT WAS DISCOVERED THAT THE MOTOR, TRIGGER FLEX, ROTOR, AND GEAR TRAIN WERE WORN.

Description of Event or Problem · 1

THE CORE UNIVERSAL DRIVER WAS RETURNED FOR SERVICE. UPON EVAL AT THE MFR, IT DISPLAYED A BIAS CURRENT ERROR WHEN CONNECTED TO THE CONSOLE, SIGNALING A CONDITION OCCURRED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN W/O USER ACTIVATION. THERE WAS NO PT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE UNIVERSAL DRIVER ERL STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK