FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 2902009 · Received November 26, 2012

Report

Report Number
1811755-2012-04159
Event Type
Malfunction
Date Received
November 26, 2012
Date of Event
October 26, 2012
Report Date
October 26, 2012
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

UPON DISASSEMBLY, IT WAS DISCOVERED THAT THE ROTOR ASSEMBLY WAS CORRODED. CORROSION CAN CAUSED INCREASED FRICTION BETWEEN ROTATING COMPONENTS, WHICH CAN CAUSE HEAT TO BE GENERATED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CORE IMPACTION DRILL HEATED UP DURING AN ORAL PROCEDURE. THERE WERE NO PT OR USER INJURIES, AND NO ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CORE IMPACTION DRILL DZJ STRYKER INSTRUMENTS KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNK