FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 2901956 · Received November 21, 2012

Report

Report Number
1824206-2012-07638
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
November 1, 2012
Report Date
November 1, 2012
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE GROUND WAS LOST BETWEEN WHERE THE POWER CORD TERMINATED INTO THE BED AND WHERE THE CIRCUIT BOARD TRAY CONNECTS TO THE UPPER FRAME. THE TECHNICIAN REMOVED A LOT OF PAINT AROUND THE SCREW HOLES THAT ATTACH THE CIRCUIT BOARD TRAY TO THE UPPER FRAME AND THIS RESOLVED THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THAT THE BED HAD LOSS OF GROUND. NO PATIENT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ADVANTA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1604

Patients

Seq Age Sex Outcome Treatment
1