FDA Adverse Event
Malfunction
Summary report: N
ADVANTA BED
MDR report key: 2901956
·
Received November 21, 2012
Report
- Report Number
- 1824206-2012-07638
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- November 1, 2012
- Report Date
- November 1, 2012
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- NURSE
Narratives
Additional Manufacturer Narrative · 1
THE TECHNICIAN FOUND THE GROUND WAS LOST BETWEEN WHERE THE POWER CORD TERMINATED INTO THE BED AND WHERE THE CIRCUIT BOARD TRAY CONNECTS TO THE UPPER FRAME. THE TECHNICIAN REMOVED A LOT OF PAINT AROUND THE SCREW HOLES THAT ATTACH THE CIRCUIT BOARD TRAY TO THE UPPER FRAME AND THIS RESOLVED THE ISSUE.
Description of Event or Problem · 1
THE TECHNICIAN ALLEGED THAT THE BED HAD LOSS OF GROUND. NO PATIENT IMPACT.