OMNIPOD INSULIN PUMP
Report
- Report Number
- 3004464228-2012-00603
- Event Type
- Malfunction
- Date Received
- November 21, 2012
- Date of Event
- October 21, 2012
- Report Date
- October 26, 2012
- Manufacturer
- INSULET CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042792
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE DEVICE WAS NOT RETURNED FOR EVAL. WE ARE UNABLE TO CONFIRM THE BENT CANNULA OR TO DETERMINE IF IT COULD HAVE CONTRIBUTED TO THE REPORTED HYPERGLYCEMIA. QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA. THE OMNIPOD'S USER GUIDE WARNS "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER". IT ADVISES, "IF YOUR BLOOD GLUCOSE IS 250 MG/DL OR ABOVE, CHECK FOR KETONES. IF KETONES ARE PRESENT, FOLLOW YOUR HEALTHCARE PROVIDER'S GUIDELINES. IF KETONES ARE NOT PRESENT, TAKE A CORRECTION BOLUS AS PRESCRIBED BY YOUR HEALTHCARE PROVIDER. CHECK BLOOD GLUCOSE REMAINS HIGH AFTER ANOTHER TWO HOURS (A TOTAL OF FOUR HOURS), REPLACE THE POD," AND "TO AVOID HYPERGLYCEMIA (HIGH BLOOD GLUCOSE) CHECK YOUR BLOOD GLUCOSE AT LEAST 4-6 TIMES A DAY (WHEN YOU WAKE UP, BEFORE EACH MEAL, AND BEFORE GOING TO BED)."
THE CUSTOMER'S MOTHER REPORTED THAT AT 9:30 AM, HER DAUGHTER'S BLOOD GLUCOSE MEASURED 392 MG/DL WITH HIGH KETONES. BOLUSES (DOSAGES AND TIMES NOT REPORTED) WERE TAKEN FOR CORRECTION, BUT HER BG REMAINED BETWEEN 386 AND 436 MG/DL. WHEN THE DEVICE WAS REMOVED (3:30 PM), THE CANNULA LOOKED SLIGHTLY BENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OMNIPOD INSULIN PUMP | PUMP, INFUSION, INSULIN | LZG | INSULET CORPORATION | 11200 | L30942 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 9 YR |