FDA Adverse Event Malfunction Summary report: N

OMNIPOD INSULIN PUMP

MDR report key: 2901867 · Received November 21, 2012

Report

Report Number
3004464228-2012-00599
Event Type
Malfunction
Date Received
November 21, 2012
Date of Event
October 25, 2012
Report Date
October 25, 2012
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR EVAL. WE ARE UNABLE TO ASSESS THE PRODUCT CONDITION. THE CUSTOMER REPORTED THAT THE CANNULA HAD DISLODGED FROM THE INFUSION SITE. THE CONDITION WOULD INTERRUPT INSULIN DELIVERY AND COULD CONTRIBUTE TO HYPERGLYCEMIA. THE OMNIPOD USER'S GUIDE WARNS "CHECK OFTEN TO MAKE SURE THE POD AND SOFT CANNULA ARE SECURELY ATTACHED AND IN PLACE. A LOOSE OR DISLODGED CANNULA MAY INTERRUPT INSULIN DELIVERY," "BECAUSE INSULIN PODS USE ONLY RAPID-ACTING INSULIN, USERS ARE AT INCREASED RISK FOR DEVELOPING HYPERGLYCEMIA (HIGH BLOOD GLUCOSE) IF INSULIN DELIVERY IS INTERRUPTED. IF IT IS UNTREATED, SEVERE HYPERGLYCEMIA CAN QUICKLY LEAD TO DIABETIC KETOACIDOSIS (DKA). DKA CAN CAUSE BREATHING DIFFICULTIES, SHOCK, COMA, OR DEATH. IF INSULIN DELIVERY IS INTERRUPTED FOR ANY REASON, YOU MAY NEED TO REPLACE THE MISSING INSULIN - USUALLY WITH AN INJECTION OF RAPID-ACTING INSULIN. ASK YOUR HEALTHCARE PROVIDER FOR INSTRUCTIONS ON HANDLING INTERRUPTED INSULIN DELIVERY," AND "TEST RESULTS GREATER THAN 250 MG/DL MEAN HIGH BLOOD GLUCOSE (HYPERGLYCEMIA). IF YOU GET RESULTS ABOVE 250 MG/DL, BUT DO NOT HAVE SYMPTOMS OF HYPERGLYCEMIA, REPEAT THE TEST. IF YOU HAVE SYMPTOMS OR CONTINUE TO GET RESULTS THAT FALL ABOVE 250 MG/DL, FOLLOW THE TREATMENT ADVICE OF YOUR HEALTHCARE PROVIDER". QUALIFICATION RECORDS WERE REVIEWED AND THE PRODUCT LOT MET ALL ACCEPTANCE CRITERIA.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE CANNULA CAME OUT OF THE INFUSION SITE AND HIS BLOOD GLUCOSE ROSE TO 395 MG/DL. HE ALSO OBSERVED THAT THE CANNULA WAS BENT. HE TOOK A MANUAL INJECTION FOR CORRECTION AND WAS ABOUT TO ACTIVATE A NEW POD AFTER THE CALL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30944

Patients

Seq Age Sex Outcome Treatment
1 68 YR