LIGACLIP*MCA SMALL APPLIER
Report
- Report Number
- 3005075853-2013-00132
- Event Type
- Malfunction
- Date Received
- January 8, 2013
- Date of Event
- December 27, 2012
- Report Date
- December 27, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDO
- PMA / PMN Number
- K820837
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? FOURTH OR 5TH. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? LOWER ABDOMINAL WALL. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? YES. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? UNK. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD?NO, ASKU. IF SO, WHEN? DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? IN. WHAT WERE THE INDICATIONS FOR SURGERY? HERNIA WHAT WAS FOUND? UNK. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? UNK. IF SO, WHAT? DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? NO. IF SO, WHOM?
(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN INGUINAL HERNIA AND LYMPH NODE BIOPSY PROCEDURE, THE DEVICE HAD BEEN FIRED FOUR OR FIVE TIMES AND THEN IT BEGAN TO EJECT CLIPS AND NOT FEED PROPERLY. THEY REMOVED THE CLIPS FROM THE PATIENT WITH GRASPERS. THEY DID USE THE DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 9525 | LIGACLIP*MCA SMALL APPLIER | CLIP, IMPLANTABLE | GDO | ETHICON ENDO-SURGERY, LLC. | NA | H44V71 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |