LIGACLIP*MCA SMALL APPLIER

MDR report key: 2901736 · Received January 8, 2013

Report

Report Number
3005075853-2013-00132
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
December 27, 2012
Report Date
December 27, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDO
PMA / PMN Number
K820837
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). WHICH FIRING OF THE DEVICE DID THIS EVENT OCCUR ON? FOURTH OR 5TH. WHAT VESSEL OR STRUCTURE WAS THE DEVICE FIRED ON AT THE TIME OF THE EVENT? LOWER ABDOMINAL WALL. WAS THE CLIP FULLY ADVANCED INTO THE JAWS PRIOR TO FIRING? YES. WAS THERE ANY TORQUING OR TWISTING OF THE DEVICE PRESENT AT THE TIME OF FIRING? UNK. WAS ANY UNEXPECTED RESISTANCE FELT WHILE FIRING THE TRIGGER? NO. WERE ANY UNEXPECTED NOISES HEARD?NO, ASKU. IF SO, WHEN? DID ANYTHING UNEXPECTED HAPPEN PRIOR TO THIS INCIDENT? NO. WAS THE DEVICE FIRED AFTER THIS INCIDENT IN OR OUT OF THE PATIENT? IN. WHAT WERE THE INDICATIONS FOR SURGERY? HERNIA WHAT WAS FOUND? UNK. DOES THE PATIENT HAVE ANY RELEVANT HISTORY OF SURGICAL TREATMENTS? UNK. IF SO, WHAT? DID SOMEBODY OTHER THAN THE PRIMARY SURGEON FIRE THE INSTRUMENT? NO. IF SO, WHOM?

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH HISTORY RECORDS WERE REVIEWED WITH NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN INGUINAL HERNIA AND LYMPH NODE BIOPSY PROCEDURE, THE DEVICE HAD BEEN FIRED FOUR OR FIVE TIMES AND THEN IT BEGAN TO EJECT CLIPS AND NOT FEED PROPERLY. THEY REMOVED THE CLIPS FROM THE PATIENT WITH GRASPERS. THEY DID USE THE DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9525 LIGACLIP*MCA SMALL APPLIER CLIP, IMPLANTABLE GDO ETHICON ENDO-SURGERY, LLC. NA H44V71

Patients

Seq Age Sex Outcome Treatment
1