FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 2901659 · Received January 2, 2013

Report

Report Number
1028232-2012-03462
Event Type
Injury
Date Received
January 2, 2013
Date of Event
November 15, 2012
Report Date
December 19, 2012
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVZ
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE ANALYSIS IS THEREFORE BASED ON THE INSPECTION OF THE QUALITY DOCUMENTS ACCOMPANYING THIS PARTICULAR DEVICE. THE MANUFACTURING PROCESS FOR THIS DEVICE WAS RE-INVESTIGATED. ALL PRODUCTION STEPS HAD BEEN PERFORMED ACCORDINGLY. THERE WAS NO SIGN OF ANY INCONSISTENCY DURING THE MANUFACTURING PROCESS, WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION. IN SUMMARY, THE DEVICE WAS NOT RETURNED FOR ANALYSIS. THE REVIEW OF THE QUALITY DOCUMENTS CONFIRMED A REGULAR DEVICE MANUFACTURING.

Description of Event or Problem · 1

OUS MDR - BOSTON SCIENTIFIC RECEIVED INFO THAT THIS RIGHT ATRIAL (RA) LEAD EXHIBITED PACING FAILURE, AS OBSERVED ON ELECTROGRAMS. AN X-RAY WAS PERFORMED, WHICH CONFIRMED THE LEAD HAD DISLODGED AND HAD FALLEN, INTO THE RIGHT VENTRICLE. THE LEAD WAS REPOSITIONED, WITH NO FURTHER ADVERSE PT EFFECTS REPORTED. THE LEAD REMAINS IN SERVICE. AS NO FURTHER INFO CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFO BE PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
681 DEXTRUS 4136 PACER LEAD NVZ BIOTRONIK SE & CO. KG 360310

Patients

Seq Age Sex Outcome Treatment
1 UNK Hospitalization