ECHELON*FLEX60 LONG
Report
- Report Number
- 3005075853-2013-00121
- Event Type
- Malfunction
- Date Received
- January 8, 2013
- Date of Event
- December 10, 2012
- Report Date
- December 17, 2012
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K081146
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. ON WHAT TISSUE TYPE WAS THE DEVICE USED? AT WHAT LOCATION ON THE TISSUE? WAS THE DEVICE FIRED ON TISSUE THAT HAD BEEN RADIATED OR HAS THE PATIENT BEEN TAKING SYSTEMIC STEROIDS? IF ECHELON, DURING WHICH STROKE DID THE EVENT OCCUR? WHAT OTHER COLOR CARTRIDGES WERE USED BEFORE AND AFTER THIS EVENT? WAS BUTTRESSING MATERIAL UTILIZED? IF SO, WHICH PRODUCT? WAS THE INSTRUMENT FIRED ACROSS OR NEAR AN EXISTING STAPLE LINE OR CLIP? WERE ANY UNEXPECTED NOISES HEARD? IF SO, WHEN? WERE ANY OF THE FORCES HIGHER OR LOWER THAN EXPECTED (CLOSING, FIRING, OR OPENING)? AFTER USE, DID EACH OF THE TRIGGERS AND BUTTONS AUTOMATICALLY RETURN TO THEIR ORIGINAL (PRE-FIRED) POSITIONS, WITHOUT INTERVENTION? IS THERE ANY OTHER ADDITIONAL INFORMATION YOU WOULD LIKE TO SHARE ABOUT THIS DEVICE OR PROCEDURE?
(B)(4). ADDITIONAL INFORMATION: THE DEVICE WAS RECEIVED FOR ANALYSIS WITH THE KNIFE ADVANCED, PREVENTING THE DEVICE FROM CLOSING. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THIS CONDITION. IN ADDITION, A RELOAD WAS RECEIVED IN A SEPARATE BAG. THE CARTRIDGE RELOAD WAS RECEIVED PARTIALLY FIRED 1/16. IT IS POSSIBLE THAT WHILE LOADING THE RELOAD, THE CARTRIDGE WAS PUSHED FARTHER BACK THAN THE CARTRIDGE ALIGNMENT STOP WINDOWS RESULTING IN THE KNIFE PUSHING THE ONE PIECE SLED FORWARD AND LOCKING THE CARTRIDGE. IT SHOULD BE NOTED THAT IN ORDER TO OPEN A DEVICE THAT HAS BEING PARTIALLY FIRED OR FULLY FIRED A REVERSE STROKE NEEDS TO BE PERFORMED TRIGGER TO TRIGGER TO HANDLE IN ORDER TO RETURN THE KNIFE TO THE HOME POSITION (INDICATOR IN THE "0" POSITION) AND PRESS THE ANVIL RELEASE BUTTON TO OPEN. THE KNIFE WAS MANUALLY RETURNED TO THE HOME POSITION AND THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH THE RETURNED RELOAD. THE DEVICE ACHIEVED ITS COMPLETE STROKE FIRING SEQUENCE WITHOUT ANY DIFFICULTIES. THE STAPLE LINE AND CUT LINE WERE COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORM SHAPE. THE EVENT COULD NOT BE CONFIRMED AS THE DEVICE CLOSED AND OPENED AS INTENDED.
(B)(4). INFORMATION IS UNAVAILABLE; DEVICE WAS NOT RETURNED FOR EVALUATION. DEVICE NOT RETURNED. THE COMPLAINT COULD NOT BE CONFIRMED BECAUSE NO DEVICE WAS RETURNED FOR ANALYSIS. THE DEVICE HISTORY RECORDS WERE REVIEWED AND THE MANUFACTURING CRITERIA WAS MET PRIOR TO THE RELEASE OF THE BATCHES INCLUDED IN THIS LOT.
IT WAS REPORTED THAT DURING A GASTRECTOMY PROCEDURE, DEVICE REMAINED CLOSED ON THE TISSUE. WITH THE FIRST CARTRIDGE (ECR60D, LOT J4C57L), THE DEVICE DID NOT STAPLE BUT REMAINED CLOSED ON THE TISSUE. THEY HAD TO USE THE RED BUTTON TO OPEN IT. THEY USED ANOTHER LIKE DEVICE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 9842 | ECHELON*FLEX60 LONG | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | J4CF67 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |