FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 2901309 · Received January 8, 2013

Report

Report Number
3005075853-2013-00119
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
December 13, 2012
Report Date
December 21, 2012
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION, THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE FINAL QUALITY RELEASE CRITERIA WAS MET BEFORE THIS BATCH WAS RELEASED FOR DISTRIBUTION.

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A CHOLECYSTECTOMY PROCEDURE, THE CLIP APPLIER MALFUNCTIONED. THE FIRST CLIP DID NOT COME OUT UNTIL THE SECOND FIRING. THE NEXT CLIP WOULD NOT ATTACH TO THE DUCT. IT WAS NOT REPORTED HOW THE PROCEDURE WAS COMPLETED. NO PATIENT IMPACT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
8587 LIGAMAX-5MM ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA J4C45K

Patients

Seq Age Sex Outcome Treatment
1