FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 2901077 · Received January 8, 2013

Report

Report Number
2531779-2013-00337
Event Type
Malfunction
Date Received
January 8, 2013
Report Date
December 10, 2012
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6). CORRECTION NUMBER: 2531779-03/24/2010-003-R.

Additional Manufacturer Narrative · 1

FOLLOW-UP #1 (B)(4) -DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: A REVIEW OF THE PUMP HISTORY INDICATED THAT LOSS OF CARTRIDGE DETECTION HAD OCCURRED WHICH WAS DUPLICATED DURING TESTING. THE PUMP WAS OPENED AND A FORCE SENSOR CIRCUIT COMPONENT WAS FOUND TO BE MISALIGNED ON THE PRINTED CIRCUIT BOARD.

Description of Event or Problem · 1

ON (B)(6) 2012, THE DISTRIBUTOR CONTACTED ANIMAS, ALLEGING LOAD STEP MALFUNCTION. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9648 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ANM 2020 INSULIN INFUSION PUMP

Patients

Seq Age Sex Outcome Treatment
1