FDA Adverse Event Malfunction Summary report: N

PROMUS ELEMENT¿ PLUS

MDR report key: 2901071 · Received January 8, 2013

Report

Report Number
2134265-2012-08349
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
December 13, 2012
Report Date
December 13, 2012
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
NIQ
PMA / PMN Number
P110010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MANUFACTURER: THE CRIMPED STENT, BALLOON AND TIP SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. NO KINKS OR DAMAGE WERE NOTICED ALONG THE SHAFT OF THE DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL PROCEDURAL FACTORS. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A PERCUTANEOUS CORONARY INTERVENTION STENT DAMAGE WAS NOTED. THE 90% STENOSED AND MODERATELY TORTUOUS TARGET LESION WAS LOCATED IN THE DISTAL RIGHT CORONARY ARTERY. DURING UNPACKING IT WAS NOTED THAT THE 3.5X32MM PROMUS ELEMENT PLUS STENT WAS DAMAGED. THE STENT WAS USED DURING THE PROCEDURE AND WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS STABLE.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING PREPARATION FOR A PERCUTANEOUS CORONARY INTERVENTION STENT DAMAGE WAS NOTED. THE 90% STENOSED AND MODERATELY TORTUOUS TARGET LESION WAS LOCATED IN THE DISTAL RIGHT CORONARY ARTERY. DURING UNPACKING IT WAS NOTED THAT THE 3.5X32MM PROMUS ELEMENT PLUS STENT WAS DAMAGED. THE STENT WAS USED DURING THE PROCEDURE AND WAS UNABLE TO CROSS THE LESION. THE PROCEDURE WAS COMPLETED WITH ANOTHER DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT CONDITION IS STABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
9647 PROMUS ELEMENT¿ PLUS STENT, CORONARY, DRUG-ELUTING NIQ BOSTON SCIENTIFIC - GALWAY H7493918432350 15255974

Patients

Seq Age Sex Outcome Treatment
1