FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 2900957 · Received December 13, 2012

Report

Report Number
1720753-2012-10409
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
December 3, 2012
Report Date
December 13, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND REPLACED THE 5 VOLT POWER SUPPLY. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM DISPLAYED A COMMUNICATION ERROR. THIS ERROR WILL CAUSE THE SYSTEM TO LOCKUP OR NOT TO BOOT. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1