FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2900937
·
Received December 13, 2012
Report
- Report Number
- 1720753-2012-10425
- Event Type
- Malfunction
- Date Received
- December 13, 2012
- Date of Event
- November 30, 2012
- Report Date
- December 13, 2012
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR BOARD AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM CONTINUED TO PRODUCE FLUOROSCOPY X-RAY FOLLOWING RELEASE OF THE FOOTSWITCH WITHIN A PATIENT PROCEDURE. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |