FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2900937 · Received December 13, 2012

Report

Report Number
1720753-2012-10425
Event Type
Malfunction
Date Received
December 13, 2012
Date of Event
November 30, 2012
Report Date
December 13, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE GENERATOR BOARD AND CONNECTORS WERE RESEATED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE. THIS MALFUNCTION MAY HAVE RESULTED IN A POSSIBLE ACCIDENTAL RADIATION OCCURRENCE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM CONTINUED TO PRODUCE FLUOROSCOPY X-RAY FOLLOWING RELEASE OF THE FOOTSWITCH WITHIN A PATIENT PROCEDURE. THERE WAS NO PATIENT INJURY OR DEATH REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1