CONSTELLATION VISION SYSTEM

MDR report key: 2900870 · Received December 12, 2012

Report

Report Number
2028159-2012-01944
Event Type
Malfunction
Date Received
December 12, 2012
Date of Event
November 12, 2012
Report Date
November 13, 2012
Manufacturer
ALCON - IRVINE TECHNOLOGY CTR
Product Code
HQC
PMA / PMN Number
K101285
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REP EXAMINED THE SYS AND CONFIRMED THE SYS MESSAGE REPORTED. THE PNEUMATICS MODULE WAS REPLACED. THE SYS WAS THEN TESTED AND MET ALL PRODUCT SPECS. THERE WAS NO SAMPLE RETURNED FOR EVAL AND NO ADD'L INFO PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADD'L SIMILAR REPORTS FOR THIS SYS. THE ROOT CAUSE CANNOT BE DETERMINED CONCLUSIVELY. (B)(4).

Description of Event or Problem · 1

A CUSTOMER REPORTED A SYS MESSAGE DISPLAYED DURING SETUP FOR A PROCEDURE. THE CASE WAS CANCELED. THE PT HAD ALREADY RECEIVED PERIBULBAR ANESTHESIA IN PREPARATION FOR THE SURGERY. THERE WAS NO PT HARM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CONSTELLATION VISION SYSTEM UNIT, PHACOFRAGMENTATION HQC ALCON - IRVINE TECHNOLOGY CTR L NA

Patients

Seq Age Sex Outcome Treatment
1