FDA Adverse Event Malfunction Summary report: N

STENOSCOP

MDR report key: 2900852 · Received December 12, 2012

Report

Report Number
1720753-2012-10350
Event Type
Malfunction
Date Received
December 12, 2012
Date of Event
November 12, 2012
Report Date
December 12, 2012
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE REPRESENTATIVE EVALUATED THE SYSTEM AND CLEANED AND RESEATED THE MEMORY MODULE. THE SYSTEM OPERATES AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM WOULD NOT COMPLETE BOOT UP. NO PATIENT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STENOSCOP FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) STENOSCOP

Patients

Seq Age Sex Outcome Treatment
1