FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2900808
·
Received December 12, 2012
Report
- Report Number
- 1720753-2012-10366
- Event Type
- Malfunction
- Date Received
- December 12, 2012
- Date of Event
- December 3, 2012
- Report Date
- December 12, 2012
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE REP EVALUATED THE SYSTEM AND COULD NOT REPRODUCE THE REPORTED PROBLEM. THE SYSTEM OPERATES AS INTENDED.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THE SYSTEM WOULD INTERMITTENTLY FAIL TO DISPLAY A VISIBLE/USABLE/STABLE FLUOROSCOPIC IMAGE ON BOOT UP. THERE IS NO REPORT OF PT INJURY OR DEATH.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |