FDA Adverse Event Malfunction Summary report: N

COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 2900628 · Received January 8, 2013

Report

Report Number
1061932-2013-00018
Event Type
Malfunction
Date Received
January 8, 2013
Date of Event
December 9, 2012
Report Date
December 9, 2012
Manufacturer
BECKMAN COULTER INC.
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED AND OBSERVED THAT THE SOURCE OF THE LEAK WAS A DISCONNECTED TUBING AT THE SHEATH RESTRICTOR. THE FSE REATTACHED THE TUBING NO FURTHER EVIDENCE OF LEAKING WAS OBSERVED. FAILURE MODE: THE CAUSE OF THE LEAK WAS A DISCONNECTED TUBING AT THE SHEATH RESTRICTOR. PER LABELING, BECKMAN COULTER, INC. URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING THAT WHILE BEGINNING STARTUP A LEAK WAS OBSERVED IN THE BACK OF THE COULTER LH 750 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS APPROXIMATELY 500 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE INSTRUMENT DID NOT GENERATE ANY FLAGS OR ERROR MESSAGES AS A RESULT OF THIS EVENT. THE CUSTOMER OBSERVED THE LEAK WHEN STARTING UP THE INSTRUMENT AT THE BEGINNING OF THE MORNING SHIFT. THE CUSTOMER WAS WEARING PERSONAL PROPER EQUIPMENT (PPE) CONSISTING OF GLOVES AND A LAB COAT AT THE TIME OF THE OCCURRENCE. THERE WAS NO CONTACT TO OPEN OR BROKEN SKIN OR MUCOUS MEMBRANES. NO ONE SOUGHT MEDICAL ATTENTION. NO SAMPLES WERE AFFECTED OR REPORTED AS A RESULT OF THE LEAK. NO ERRONEOUS RESULTS WERE GENERATED.